EDAN F6 User manual - page 152
F Series Fetal & Maternal Monitor User Manual Product Specifications
- 141 -
I
sppa.3
2
I
spta.3
2
Max Output Power
Effective Radiating Area:
942 mm
2
± 15%
FHR Measurement Range:
50 bpm ~ 240 bpm
Resolution: 1
bpm
Accuracy:
±2 bpm (F6/F6 Express)
±1 bpm (F9/F9 Express)
Dielectric Strength:
4000Vrms
Maximum transducer
temperature rise during
use:
Less than 5
qC (
9ºF
)
Expanded uncertainty of
temperature test:
U=0.4
qC (
0.72ºF
), k=2
DECG
Technique:
Peak-peak detection technique
DFHR Measurement
Range:
30bpm ~ 240bpm
Resolution: 1bpm
Accuracy: ±1bpm
Input Impedance:
> 10M (Differential, DC50/60Hz)
Input Impedance:
> 20M (Common Mode)
CMRR: >
110dB
Noise:
4μVp
Skin Voltage Tolerance:
±500mV
Fetal Input Voltage
Current:
20μVp-3mVp
TOCO
TOCO Range:
0% ~ 100%
Non-linear Error:
±
10%
Baseline Drift due to
Temperature Changes
1 unit/min/°C (free air)
5 units/min/°C (underwater)
Resolution: 1%
Zero Mode:
Automatic(TOCO value becomes zero or below
lasting for 30 seconds)/ Manual
Dielectric Strength:
4000Vrms
IUP
Pressure Range:
0 ~ 100mmHg
Sensitivity: 5μV/V/mmHg