EDAN F6 User manual - page 50
F Series Fetal & Maternal Monitor User Manual Alarms
- 39 -
SpO
2
Alarm Delay
0 second, not adjustable
0 second
SpO
2
Alarm Level
Medium, not adjustable
Medium
SYS Alarm
On, Off
On
SYS Low Alarm Limit
40 mmHg ~ 269 mmHg, in increments of 1
90 mmHg
SYS High Alarm Limit
41 mmHg ~ 270 mmHg, in increments of 1
160 mmHg
SYS Alarm Delay
0 second, not adjustable
0 second
SYS Alarm Level
Medium, not adjustable
Medium
DIA Alarm
On, Off
On
DIA Low Alarm Limit
10 mmHg ~ 214 mmHg, in increments of 1
50 mmHg
DIA High Alarm Limit
11 mmHg ~ 215 mmHg, in increments of 1
90 mmHg
DIA Alarm Delay
0 second, not adjustable
0 second
DIA Alarm Level
Medium, not adjustable
Medium
TEMP Alarm
On, Off
On
TEMP Low Alarm Limit
0 ºC ~ +49.9 ºC, in increments of 0.1
+36.0 ºC
TEMP High Alarm Limit
+0.1 ºC ~ +50.0 ºC, in increments of 0.1
+39.0 ºC
TEMP Alarm Delay
0 second, not adjustable
0 second
TEMP Alarm Level
Medium, not adjustable
Medium
NOTE:
The upper limit must be higher than the lower limit. When setting the upper limit, you do
not have access to the options that are lower than the preset lower limit, and vice versa.