EDAN F6 User manual - A1.4 Recorder Specifications
F Series Fetal & Maternal Monitor User Manual Product Specifications
- 144 -
Infrared light
(905
r10) nm
Emitted light energy
Information about the wave length range can be especially useful to clinicians
(for instance, when photodynamic therapy is performed.)
NIBP
(for adult)
Measurement Range:
Systolic pressure:
40mmHg ~ 270mmHg
Diastolic pressure:
10mmHg ~ 215mmHg
Measuring Time:
120 seconds
Software Over Voltage
Protection:
(297
r 3) mmHg
Hardware Over Voltage
Protection:
(320
r 10) mmHg
Cuff pressure measuring
range:
0 mmHg ~ 300 mmHg
Resolution: 1
mmHg
Measuring Accuracy
Max. average deviation:
r5mmHg
Max. standard deviation:
8mmHg
TEMP
Channel: 1
Measuring Mode:
Direct Mode
Position: Axilla
Measurement Range:
0ºC ~ +50ºC (+32ºF ~ +122ºF)
Accuracy:
(sensor error excluded)
0ºC ~ +25ºC (+32ºF ~ +77ºF):
± 0.2ºC (±0.36ºF)
+25ºC ~ +45ºC (+77ºF ~ +113ºF): ± 0.1ºC (±0.18ºF)
+45ºC ~ +50ºC (+113ºF ~ +122ºF): ± 0.2ºC (±0.36ºF)
Refresh Time:
Every 1s to 2s
Accessory: TEMP
sensor
A1.4 Recorder Specifications
Paper:
Z-fold, thermosensitive
(compatible with GE and PHILIPS recorder paper)
Paper width:
152mm (GE), 150mm (PHILIPS)