EDAN F6 User manual - 4.11 Testing Alarms
F Series Fetal & Maternal Monitor User Manual Alarms
- 38 -
4.11 Testing Alarms
To test the functions of visible and audible alarms, do the following:
1 Switch on the monitor.
2 Enable the alarm.
3 Set the alarm limits to a small range.
4 Stimulate a signal that is higher than the upper limit or lower than the lower limit. Or
disconnect one of the plugs.
5 Verify if the visible and audible alarms are working properly.
4.12 Patient Alarm Defaults
Alarm Setting
Options
Default
High Level
ASYSTOLE
On (not adjustable)
On
Asystole Alarm Delay
0 second (not adjustable)
0 second
Asystole Alarm Level
High (not adjustable)
High
Medium Level
FHR1/FHR2 Alarm
On, Off
On
FHR1/FHR2 Low Alarm Limit
60 bpm ~ 205 bpm, in increments of 5
110 bpm
FHR1/FHR2 High Alarm Limit 65 bpm ~ 210 bpm, in increments of 5
160 bpm
FHR1/FHR2 Alarm Delay
0 ~ 30
second(s), in increments of 5
10 seconds
FHR1/FHR2 Alarm Level
Medium, not adjustable
Medium
HR Alarm
On, Off
On
HR Low Alarm Limit
30 bpm ~ 239 bpm, in increments of 1
50 bpm
HR High Alarm Limit
31 bpm ~ 240 bpm, in increments of 1
120 bpm
HR Alarm Delay
0 second, not adjustable
0 second
HR Alarm Level
Medium, not adjustable
Medium
SpO
2
Alarm
On, Off
On
SpO
2
Low Alarm Limit
50% ~ 99%, in increments of 1
90%
SpO
2
High Alarm Limit
51% ~ 100%, in increments of 1
100%