Zeiss HUMPHREY 720i User Manual - page 181
Manual is about: FIELD ANALYZER II - i series
of
328
DATABASE MANAGEMENT
9-15
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER II -
i series
5
If recovery is unsuccessful, a
user message appears.
Select OK. Go to next option:
“Rebuild Hard Disk Database”.
Hard drive failure:
In the event the FIX NOW recovery attempts have failed, use the rebuild utility (REBUILD HARD
REBUILD HARD DISK
DISK DATABASE). Typically a lengthy process, rebuilding the database frequently corrects the
DATABASE
original problem and avoids the inconvenience of restoring the database from backup floppy
disks or magneto-optical disks.
1
From the Main Menu, press
the SYSTEM SETUP icon.
9.htp
3/14/03, 3:19 PM
15