If the rebuild utility is
unsuccessful, the screen
displays a user message.
Choose the desired button.
RETRY repeats the rebuild process. Occasionally, a second try is successful.
DELETE initializes the database. All data on the Hard Disk will be deleted. You will need
to restore the database from magneto-optical or floppy disks as described on the following
pages.
CANCEL attempts to return the HFA II to its previous condition. Use the “Floppy Only” method
of data storage until you find a more convenient time to deal with the failed database. When
time permits, you will need to restore the database from magneto-optical or floppy disks as
described on the following pages.
If all previously cited methods to correct the database problem have failed and you are faced
with an unrecoverable database, the only option remaining is to restore the hard disk database
from backup floppy disks or magneto-optical disks. Detailed instructions follow. If you need
assistance, call Carl Zeiss Meditec Customer Service at 1-877-486-7473 (1-877-HUMPHREY).