Zeiss HUMPHREY 720i User Manual - page 279
Manual is about: FIELD ANALYZER II - i series
of
328
CARE AND CLEANING
12-9
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER II -
i
series
REPLACING FUSES
1
Turn off the power. Unplug the
power cord that is attached to the
table at the base of the lifting column.
Move the table away from the wall to
access the back of the table.
2
Use needle-nose pliers or
fingernails to gently slide the fuse
drawer from its housing.
3
Remove defective fuses. Insert
new fuses in housing.
4
Replace the fuse drawer. Plug in
the power cord.
Note: For continued protection against fire hazard, replace only with same type and
rating of fuse.
12.htp
3/14/03, 2:45 PM
9