Zeiss HUMPHREY 720i User Manual - page 224
11-16
KINETIC TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER II -
i series
4
Press AUTOMATIC.
You will then need to select the
eye to test. The Patient Data
screen appears. Press the
PROCEED button when all desired
patient data has been entered.
5
Press START to begin the
automatic test. The meridian
to be tested is indicated on the
screen.
6
The stimulus will light but
will not move for a short
period of time. You will then see a
small diamond begin to move
across the screen and the STOP
STIMULUS button will become
active.
When the patient presses the response button, the isopter letter will appear and the next
meridian to be tested is displayed on the screen. The HFA II will delay, then start to move the
stimulus. This is the routine throughout the entire test.
the diamond
11.htp
3/13/03, 3:01 PM
16