Zeiss HUMPHREY 720i User Manual - page 200
10-12
CUSTOM TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER II -
i
series
Erasing grids
or single points
Please read this entire section before removing any points. Unwanted Grids or Single Points
can be removed from a Custom test that you are creating. However, only the last entry can be
removed. If you input a Single Point or Grid that you are unsure of, it is best to erase it and
input it again later. The ERASE LAST ENTRY button is disabled if there are no points.
1
From the Custom screen,
press ERASE LAST ENTRY.
The last Single Point or Grid you
have entered will automatically
be removed from the test field.
No warning will be issued
before deleting the last entry.
2
Subsequent selections of the ERASE LAST ENTRY button will delete the next previous entry.
When all desired test points have been entered, you must save the test. Saving the Custom test
pattern you programmed creates a test button on the Custom Test screen which is specific to
that test. This button is used to recall the test pattern for future use. Be sure to save your
Custom test pattern as described below.
1
From the Custom screen,
select SAVE TEST.
Saving custom tests
10.htp
3/14/03, 2:43 PM
12