Zeiss HUMPHREY 720i User Manual - page 229
11-21
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER II -
i series
KINETIC TESTING
Retesting a meridian
1
A meridian can be retested
when using the Step-by-Step
procedure. Press PREVIOUS
MERIDIAN.
2
The previous meridian will
be displayed at the top of the
screen. Press TEST MERIDIAN to
start the stimulus.
3
A second capital letter will be
displayed on the retested
meridian.
Note: When isopter lines are
drawn, the line will connect to
the second stimulus mark.
When the last point of an isopter is tested, the change of isopters may be subtle. You may note
the change by looking at the Stimulus Value at the top of the screen. You may briefly see the
announcement: “Please Wait. Preparing the Instrument For Test” as previously shown in the
Automatic mode description.