Jablotron JA-63 „PROFI“ Installation Manual - page 17
JA-63 „PROFI“ Alarm System
- 17 -
MGK51404
10.29 New service code setting
sequence: 5 nSC nSC
The Service Code can be used to enter the programming mode. A new Service Code must be entered twice in a row to
avoid an error.
To change the code enter:
5 nSCnSC
where nSC is your new Service Code (four digits)
Example: to change service code to 1276 enter: 5 1276 1276
Factory default setting: service code is 6060
10.30 User Mode entering
Sequence: 6 9 9 9
This sequence is used to switch from the Programming Mode to the User Mode, where you can set zones´ bypass (see
User’s manual). You can exit the User Mode by pressing the “N” button. The bypassed zones will remain active after the
leaving the User Mode.
10.31 Real time and date setting
sequence: 4 hh mm dd MM YY
The control panel has a built in real time clock. All events are stored to the event memory including the time of the
event. The clock should be set after the installation is completed. Time Setting:
4 hh mm dd MM YY
where hh = hours (24 hr. cycle)
mm = minutes
dd = day
MM = month
RR = year
Example: on Jun. 30 2007 at 17:15 enter: 4 17 15 30 06 07
After the control panel is powered, its internal clock’s default setting is: 00 00 01 01 00
Note: detail control panel event history can be viewed with a connected PC using Comlink software.
11 System testing
For testing by installer, the control panel should be in the programming mode - "P" indicated on the keypad’s LED (F0
Service Code). Testing can also be done by a user in the user mode (confirmed by “U”). The user mode is accessible
with the Master code. To open the user mode enter F 0 Master Code when the control panel is disarmed.
No alarm can be triggered in programming or user modes and any triggering of a detector (wireless or wired) will result
in a “beep” (press F to select a loud “beep” generated by a wired siren) and the display will briefly show which zone was
triggered. Enrolled wireless controllers, sirens and other items’ signals will be similarly indicated.
• Some detectors (JA-60P, JA-60N, JA-60B etc.) have an extra testing mode, which is activated for 5 minutes
after the detector’s cover is attached (see manuals of the particular detectors). If the detector is in testing mode,
it will indicate triggering locally with its LED, and it will also indicate the triggering on the control panel keypad’s
LED. Note that the JA-60P motion detector in normal mode (after 5 minutes testing mode) can not send next
triggering information until 5 minutes after the previous triggering was sent (this period can be shortened to 1
minute - see setting of the JA-60P detector).
• Triggering of a detector wired to one of the L1 to L4 inputs is indicated on the control panel keypad’s LED for
about 2 seconds after the triggering. So, if a detector is permanently triggered for a longer period, it will not be
indicated. If a double balanced input loop (2x 1k1) is used, then the control panel distinguishes triggering of the
detector from tampering.
• The best way of testing is via a connected PC using the Comlink software (see section 17). In the service
events window you will see a chronological record of all performed tests, including zone setting, quality of
communication etc.