Abbott i-STAT 1 System Manual - Quality Control
QUALITY CONTROL
14
Art: 714376-00M
Rev. Date: 31-Jan-12
14-1
Overview
This section describes the steps to be taken to verify the performance of the
analyzer and cartridges. The rationale for the i-STAT cartridge and analyzer
quality regimen is described in the Theory section of this manual.
Customization
The quality control behavior of the analyzer can be customized via the Central
Data Station or analyzer Customization Change function to:
• turntheexternalElectronicSimulatorreminderonoroff.
• prompttheoperatortousetheexternalElectronicSimulatorat
scheduled intervals.
• turntheinternalElectronicSimulatoronoroff,andselectthe
simulator test cycle intervals.
• disablefurthercartridgetestingwhentheinternalElectronic
Simulator test fails.
See the Customization section for default values.
Data Retention
Quality control data is transmitted to the Central Data Station. If a Central
DataStationisnotbeingused,thechartsattheendifthissectioncanbeused
to record liquid and electronic control results.
QUALITY CONTROL FOR i-STAT CARTRIDGES AND THE ANALYZER’S CARTRIDGE TEST CYCLE
Verify Performance of
Analyzers Daily
VerifytheperformanceofeachanalyzeronsiteusingtheElectronicSimulator
(externalorinternal)onceadayonthedaystheanalyzersareinuse.Notethat
regulatory or accreditation requirements may dictate more frequent intervals.
Verify Newly Received
Cartridges
1. Verifythatthetransittemperaturesweresatisfactoryusingthefour-
windowtemperatureindicatorstripincludedintheshippingcontainer.
2. Fromeachlotineachshipmentofcartridges,analyzemultiplelevelsof
i-STATcontrols(andRNA
®
Medicalcontrolsiftestingforhematocrit)
using any verified analyzer.*
* This is not a manufacturer's system instruction; it is a suggestion to
complywithregulatoryrequirements.
Check Refrigerator
Storage Daily
Verifythatthecartridgesstoredintherefrigeratorarewithintheexpiration
dateprintedontheboxes.
Verifythatthestoragerefrigeratordidnotexceedthetemperaturelimitsof2to
8°C(35to46°F).Ifstorageconditionsareindoubt,usecontrolstoverifythat
the cartridges are performing properly. This is especially important if freezing
conditions are suspected at the back of the refrigerator.