Abbott i-STAT 1 System Manual - page 329
Art: 714383-00F
Rev. Date: 02-Sep-08
21-5
2. Save the settings by selecting
9 (Save and Exit) at the Your choice ?
prompt.
3. Remove power and connect the Downloader in its intended location.
Troubleshooting
If a wrong number is entered, which cannot be corrected, press the
Enter key
until the session is completed and start from the beginning again.
Configure for
i-STAT 1 Serial
Downloader
Connection
Channel 1 provides network access for the i-STAT 1 Analyzer data transmissions
to a data manager. This section describes how to set up parameters for Channel
1.
1. Determine the following information:
• The IP Address of the data manager. (Example: 10.10.12.184)
• The service port number set to receive transmissions from the i-
STAT 1 Analyzer (default 6004).
2. At the
Your choice ? prompt, Select 1 (Channel 1 Configuration).
3. At each of the prompts enter the following bold-faced value:
Note: If the information to be entered is the same as the default
value, press the Enter key.
• Baudrate (38400) ?
38400 (must be set to 38400)
• I/F Mode (4C) ? (press
• Flow (00) ? (press
• Port No (10001) ? (press
• ConnectMode (C1) ?
C1 (must be set to C1)
• Auto increment source port (N) ?
N (press
• Remote IP Address : (000)
10.(000)10.(000)12.(000) 184
Note: Set Remote IP Address to the IP Address of the
computer where the data manager resides.
• Remote Port (06004) ?
6004
Note: Remote Port refers to i-STAT 1 service port defined in
the data manager.
• DisConnMode (00) ? (press
• FlushMode (44) ?
44(must be set to “44”)
• DisConnTime (00:30) ?
00:30 (Disconnect time must be 30 sec-
onds)
• SendChar 1 (00) ? (press
• SendChar 2 (00) ? (press