Abbott i-STAT 1 System Manual - page 377
Rev. Date: 07/18/06
Art: 714384-00D
22-43
CustomizAtion WorKsPACe
overview
This workspace is used to create profiles with site specific test characteristics
for the analyzers. See the Customization section of this manual for details of
items that can be customized and their default settings.
enabling
Customization
To enable customization, click the box to check it. When customization is
enabled, the Central Data Station will check the Customization Profile for
the location each time an analyzer is downloaded. If the location has the
enable updates option checked, the Central Data Station will update the
analyzer with the current Customization Profile for that location as noted below.
•
Analyzers designated to Report location as download location in the
Instrument workspace will be updated with the Customization Profile
assigned to the download location, regardless of the location to which
the instrument is assigned. Care should be taken when downloading
instruments from locations other then their assigned location.
•
Analyzers designated to Always report location as this assignment in the
Instrument workspace will always be updated with the Customization
Profile for the instrument’s assigned location, regardless of the physical
location from which it downloads.
If a location has the
enable updates option unchecked, downloading
from that location will result in no customization changes being made to
analyzers designated to
report location as download location. Analyzers
designated to
Always report location as this assignment will not be updated
if the assignment location for the instrument has the
enable updates option
unchecked, regardless of the setting associated with the physical location from
which it downloads.
User can also disable/enable CLEW updates by location. The default setting is
to have the CLEW updates occur automatically for all locations. To disable a
particular location, simply click on the corresponding check box under
update
CLeW to remove the check mark.
Password
The Customization Workspace is password protected. If the Security feature is
disabled, the default password is
istat. If the Security feature is enabled, the
user uses the same password as their CDS application logon password. To
change the password, select
tools and Change Password from the menu bar.
A password from 3 to 8 characters can be used.