D.8.15 Calibrating the Optional Temperature Probe
The optional temperature probe adjustment is made on the Sensor Interface board.
With the towfish open, turn the towfish on, and let the system boot up. Start
SonarPro. With the towfish at ambient temperature, measure the ambient
temperature with an accurate thermometer. Set the potentiometer on the Sensor
Interface board, which is shown in Figure D-26, to the current temperature. The
temperature can be viewed in the Information window in SonarPro.
D.8.16 Towfish Power Supply Pinouts
The towfish power supply pinouts are shown in Figure D-27 .