Mahr Digimar 817 CLM Operating Instructions Manual - page 136
136
Mahr GmbH • Digimar 817 CLM
Adjust the nomimal values
3. Create new program
4. Change existing program
5. Print out current program
1. <-_______
2. Change program head
3. Change program form
4. Change record head
5. Save changes
Please refer to chapter 7.3.3 Create program steps
Step
:
1
Charact
:
Group n
o
:
0
Group
: Main group
Function
:
xxxx
Comment
:
Nom. val.
:
20.012
Channel
: 0
Upp. all.
:
0.000
UCL X
: 0.000
Low. all.
:
0.000
LCL X
: 0.000
Paret. UT
:
1.000
LCL RIS
: 0.000
Paret. LT
:
1.000
UCL RIS
: 0.000
Step
:
1
Charact
:
Group n
o
:
0
Group
: Main group
Function
:
xxxx
Comment
:
Nom. val.
:
20.000
Channel
: 0
Upp. all.
:
0.000
UCL X
: 0.000
Low. all.
:
0.000
LCL X
: 0.000
Paret. UT
:
1.000
LCL RIS
: 0.000
Paret. LT
:
1.000
UCL RIS
: 0.000
Enter the „nominal values“ of the gage blocks. Skip
with the arrow key to the value 20.012 and adjust.
Confirm with the „ON/OFF“ key, adjust all the nominal
values, after the last characteristic press the „Abort“
key, to exit this menu and to save the adjusted values.
ON
OFF
1. <-_______
2. Change program head
3. Change program form
4. Change record head
5. Save changes
Use the following keys to move within the input
box:
+ / -
to scroll between program steps
Left and right arrow keys to move within the
input box
Tab function - to skip between input boxes
Skip to the start (Step__ 1)
Switch between upper and lower case