Mahr Digimar 817 CLM Operating Instructions Manual - page 37
37
Mahr GmbH • Digimar 817 CLM
4.1.2 Calibrate a twin probe
–
Press the „calibration“ key
–
Press the function key „Twin probe“.
The probe automatically travels the center of the groove of the
setting standard
–
Slide the setting standard, so that the probe is inside the groove.
The groove will automatically be measured twice
–
Position the vertical probe above the ledge
–
Press the function key „Calibrate vertical probe“.
The measured value is automatically accepted
–
Position the horizontal probe above the ledge
–
Press the function key „Calibrate horizontal probe“.
The measured value is automatically accepted
–
The probe constant of the horiziontal probe and the distance
between the horiziontal and vertical probes will be displayed.
The calibration procedure is complete
–
Select horizontal or vertical probe
Description / Sequence
Symbols / Pictures