Measuring with a taper probe
Area of application:
For rapid determination of coordinates on workpieces with a rough
tolerances (e.g. master gage for holes, perforated plate, threads. . .
+/- 0.2 mm
Procedure:
–
Measure a bore with a taper probe.
Requirement: the taper probe is calibrated. The bore must be
smaller the diameter of the taper <30.00 mm
–
Note the measured center of the bore e.g. 207.516 mm
Description / Sequence
Symbols / Pictures
–
Loosen the knurled screw and remove the standard probe,
replace
this with a taper probe and
tighten up the knurled screw.
–
Calibrate the probe
–
Select taper probe
–
Enter the diemater of the taper
–
Confirm by pressing the the taper symbol