EDAN iM80 User Manual - page 104
Patient Monitor User Manual Monitoring CO
2
- 93 -
Connecting Sensor
4
To zero the sensor, please refer to zeroing the sensor.
5 Install the airway adapter at the proximal end of the circuit between the elbow and the
ventilator Y-section.
Connecting Airway Adapter
WARNING
1
No routine user calibration required.
2
Accuracy is affected by temperature and barometric pressure.
3
It is forbidden to insert or draw out the module when the monitor is working, for it can
cause instability of the system. If you do it inadvertently, please turn off the module in
menu immediately. The module enters STANDBY mode if you reconnect it to monitor
which it is powered on. If the readings are inaccurate, you should do calibration.
NOTE:
1
Replace the airway adapter, if excessive moisture or secretions are observed in the
tubing or if the CO
2
waveform changes unexpectedly without a change in patient