NIBP Signal Sturated
Signal amplitude too strong
Low
Stop the patient from
moving.
NIBP Init Pressure
High
The initial pressure is too
high during measuring
High
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify
biomedical
engineer or
manufacturer’s
service staff.
NIBP Invalid Reset
The hardware pressure is
too high
Low
Measure again, if
failure persists, stop
using measuring
function of NIBP
module and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Time Out
Measuring time has
exceeded the specified time.
Low
Measure again or use
other measuring
method.
NIBP Tube Leak
NIBP cuff or pump has a
leakage.
Low
Check the NIBP cuff
and pump for
leakages.
NIBP Cuff Type Error
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient
type and change the
cuff.
Air Pressure Error
Environment atmospheric
pressure abnormal or
system pressure abnormal.
Low
Check whether the
airway is occluded or
pressure sensor
works properly in
pressure meter mode.
If the problem still
exists, contact your
service personnel.
NIBP Self Test Error
Sensor or other hardware
errors.
High
Contact your service
personnel.