EDAN iM80 User Manual - A Product Specification
Patient Monitor User Manual Product Specification
- 143 -
A Product Specification
A.1 Classification
Anti-electroshock type
Class I equipment and internal powered equipment
EMC type
Group 1, Class A
Anti-electroshock degree
ECG
(RESP), TEMP, IBP, C.O., Quick Temp CF
SpO
2
, NIBP, CO
2
, AG
BF
Ingress Protection
IPX1 (No protection against ingress of water if configured
with Quick TEMP module)
Disinfection/sterilization method
Refer to Chapter Care and Cleaning for details.
Working system
Continuous operation equipment
Compliant with Standards
IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+A2,
IEC/EN 60601-1-2:2001+A1, ISO 9919, ISO 21647,
IEC/EN 60601-2-27,
IEC/EN 60601-2-30, IEC/EN
60601-2-34, IEC/EN 60601-2-49, ANSI/AAMI SP10,
IEC/EN 60601-2-25, AAMI/ANSI EC13, EN12470-4
EN1060-1 EN1060-3, EN1060-4
A.2 Physical Specifications
A.2.1 Size and Weight
Product
Size
Weight
iM80
370 mm (L) × 175 mm (W)× 320 mm (H) 7 kg
iM50
260 mm (L) × 140 mm (W)× 205 mm (H) 3.6kg
A.3 Environment Specification
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
Temperature
Working
+5
°C ~ +40°C
Transport and Storage
-20
°C ~ +55°C
Humidity
Working
25% ~ 80% (non-condensing)
Transport and Storage
25% ~ 93% (non-condensing)