EDAN iM80 User Manual - page 99
Patient Monitor User Manual Monitoring IBP
- 88 -
9. Press the Start button, the device will begin calibrating.
10. Wait for the calibrated result. You should take corresponding measures based on the prompt
information.
11. After calibration, disassemble the blood pressure tubing and the attached 3-way valve.
1: Hydrargyrum pressure meter; 2: 3-way connector; 3: 3-way stopcock; 4: Pressure transducer; 5:
Pressure transducer interface cable; 6: Monitor
IBP Calibration
15.9 Troubleshooting the Pressure Calibration
The status line lists the probable reasons of an unsuccessful calibration.
Message
Corrective Action
Sensor Off, Fail!
Make sure that sensor is not off, then start the calibration.
Contact service technician if necessary.
Unable to calibrate in Demo
Mode.
Make sure that the monitor is not in DEMO mode. Contact
service technician if necessary.
Pressure out of normal range,
fail.
Make sure that you have selected transducer value in IBP
CAL, then start the calibration. Contact service technician if
necessary.
Pulsate Pressure Calibration Fail Make sure that pressure value shown by hydrargyrum
pressure meter is changeless. Contact service technician if
necessary.
15.10 IBP Alarm
When Alarm Switch is set to On, the physiology alarm occurs if any measurement value of
Systolic pressure, Mean pressure, or Diastolic pressure exceeds alarm limit. Users can adjust the
alarm limit by accessing XX Options (XX stands for the label name) > Setup > SYS Alarm /
MAP Alarm / DIA Alarm.