EDAN iM80 User Manual - page 66
Patient Monitor User Manual Monitoring ECG
- 55 -
Select sites with intact skin, without impairment of any kind.
Shave hair from sites, if necessary.
Wash sites thoroughly with soap and water. (Never use ether or pure alcohol, because this
increases skin impedance).
Rub the skin briskly to increase capillary blood flow in the tissues and remove skin scurf and
grease.
8.6.2 Connecting ECG Cables
1. Attach clip or snap to electrodes prior to placement.
2. Put the electrodes on the patient. Before attaching, apply some conductive jelly on the
electrodes if the electrodes are not electrolyte self-supplied.
3. Connect the electrode lead to the patient's cable.
CAUTION
To protect the monitor from damage during defibrillation, for accurate ECG information
and to protect against noise and other interference, use only ECG electrodes and cables
specified by EDAN.
8.7 Selecting Lead Type
To change the lead type, please:
1. Select the ECG parameter area, open the ECG Setup menu;
2. Set Lead Type to 3 Leads, 5 Leads or 12 Leads based on the lead used.
8.8 Installing Electrodes
NOTE:
The following table gives the corresponding lead names used in Europe and America
respectively. (Lead names are represented by R, L, F, N, C, C1-C6 in Europe, whose
corresponding lead names in America are RA, LA, LL, RL, V, V1-V6.)
AHA (American Standard)
IEC (Europe Standard)
Electrode Labels
Color
Electrode Labels
Color
RA
White
R
Red
LA
Black
L
Yellow
LL
Red
F
Green
RL
Green
N
Black
V
Brown
C
White