EDAN iM80 User Manual - page 139
Patient Monitor User Manual Using Battery
- 128 -
WARNING
5
Do not use, leave battery close to fire or other places where temperature may be
above 60
°C.
6
Do not immerse, throw, or wet battery in water/seawater.
7
Do not destroy the battery: do not pierce battery with a sharp object such as a needle;
Do not hit with a hammer, step on or throw or drop to cause strong shock; Do not
disassemble or modify the battery.
8
Use the battery only in the monitor. Do not connect battery directly to an electric outlet
or cigarette lighter charger.
9
Do not solder the leading wire and the battery terminal directly.
10
If liquid leaking from the battery gets into your eyes, do not rub your eyes. Wash them
well with clean water and go to see a doctor immediately. If liquid leaks of the battery
splash onto your skin or clothes, wash well with fresh water immediately.
11
Keep away from fire immediately when leakage or foul odor is detected.
12
Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
13
Do not use a battery with serious scar or deformation.
24.4 Replacing the Battery
To install or replace the battery, please follow the procedure:
1. Pull the battery door downwards to open it accoding to indication on it.
2. Pull the metal retainer until the battery can be removed.
3. Insert the new battery into the battery compartment.
4. Pull the metal retainer downward to fix the battery and close the battery door.
Battery Door