EDAN iM80 User Manual - page 26
Patient Monitor User Manual Basic Operation
- 15 -
1
SD Card
2
Nurse call port, this port is connected to the call system. When
there is an alarm, the monitor outputs nurse call signal to notify
the nurse.
3
Defibrillator synchronization / analog output. When the user
selects Analog Output, the monitor outputs the waveform
through the auxillary output port. When the user selects
Defibrillator, the monitor outputs the defibrillator
synchronization signal through the auxillary output port.
4
VGA output
5
USB interface, this port is used to connect the USB device.
6
RS232 interface
7
Network interface, this port is used to connect to the central
monitoring system through the standard network wire.
8
Anti-theft lock interface
9
Heat emission hole
10
Security lock, used to prevent the power supply cord from falling.
11
Equipotential grounding terminal, if the monitor or other
processing unit are used in internal examinations on the heart,
ensure that the room incorporates an equipotential grounding
system to which the monitor and other processing unit have
separate connection.
12
Speaker
13
Fan
3.1.3 iM50/iM80 Configuration
iM50/iM80 configuration is listed as below:
Model
Size (L×W×H)
Figure
Function
Configuration
iM50
260 mm (L) × 140 mm (W) × 205 mm (H) Square
ECG, RESP, SpO
2
,
NIBP, IBP ,TEMP,
Quick TEMP , CO
2
iM80
370 mm (L) × 175 mm (W) × 320 mm (H) Square
ECG, RESP, SpO
2
,
NIBP, TEMP, IBP,
C.O., CO
2
, AG
3.2 Operating and Navigating
Everything you need to operate the monitor is contained on its screen. Almost every element on
the screen is interactive. Screen elements include measurement numerics, waveforms, screen keys,