Patient Monitor User Manual Monitoring NIBP
- 79 -
The system will automatically inflate the pneumatic system to about 180 mmHg.
After 20 seconds, the system will automatically open the deflating valve, which marks
the completion of a pneumatic measurement.
If the prompt of Leak Test OK appears, it indicates that the airway is in good situation
and no air leaks exist. However if the alarm information of NIBP Cuff Leak appears,
it indicates that the airway may have air leaks. In this case, the user should check for
loose connection. After confirming secure connections, the user should re-perform the
pneumatic test. If the failure prompt still appears, please contact the manufacturer for
repair.