EDAN iM80 User Manual - page 171
Patient Monitor User Manual Product Specification
- 160 -
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
5
5
5
80
50
15
41 – 70 mmHg: ± 2.5% additional error
71 – 100 mmHg: ± 4% additional error
101 – 150 mmHg: ± 5% additional error
*Additional worst case error when compensation
for P
B
, O
2
, N
2
O, anesthetic agents, or helium is
correctly selected for the actual fractional gas
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38mmHg.
Barometric Pressure on EtCO
2
Measurement Values:
Quantitative effect
Ambient Barometric, Operational
0 – 40 mmHg: ± 1 mmHg additional error
41 – 70 mmHg: ± 2.5% additional error
71 – 100 mmHg: ± 4% additional error
101 – 150 mmHg: ± 5% additional error
*Additional worst case error when compensation for P
B
, O
2
, N
2
O, anesthetic agents, or helium
is correctly selected for the actual fractional gas constituents present.
A.16 C.O.
Method
Thermodilution Technique
Measuring range
C.O.
0.1 L/min ~ 20L/min
TB
23
°C ~ 43°C(73.4 º F ~109.4 º F)
TI
-1
°C ~ 27°C(30.2 º F ~80.6 º F)
Resolution
C.O.
0.1L/min
TB, TI
+0.1
°C (+0.1 º F)
Accuracy
C.O.
For C.O.
≥ 4.0 L/min: ±5%