EDAN iM80 User Manual - page 168
Patient Monitor User Manual Product Specification
- 157 -
A.11.2 SpO
2
from Nellcor Module
Measuring Range
1% ~ 100%
Resolution
1%
Accuracy
Sensor Type
Accuracy
MAX-A,
MAX-AL,
MAX-N,
MAX-P,
MAX-I, MAX-FAST
± 2 (70% ~ 100% SpO
2
)
OxiCliq A, OxiCliq P,
OxiCliq N
(Adult),
OxiCliq N (Neonate),
OxiCliq I
± 2.5 (70% ~ 100% SpO
2
)
D-YS (Infant to Adult),
DS-100A,
OXI-A/N,
OXI-P/I
± 3(70% ~ 100% SpO
2
)
D-YS (including D-YSE
ear clip), D-YS
(including
D-YSPD
spotclip)
± 3.5(70% ~ 100% SpO
2
)
* When the sensor is used to neotate as recommendation, the specified accuracy range of the
neotate is always higher ±1 than adult.
Pulse Rate
Measuring Range
20bpm ~ 300bpm
Resolution
1bpm
Accuracy
± 3bpm (20bpm ~ 250bpm)
Sensor
Wave length: approximately 660 and 900nm
Emitted light energy: ≤15mW
A.12 TEMP
Channel
2
Sensor type
YSI-10K and YSI-2.252K
Measuring Range
0
°C ~ 50 °C(32 º F ~ 122 º F)
Resolution
0.1
°C (0.1 º F)
Accuracy (Without sensor)
±0.1°C or ±0.2 º F
Refresh Time
Every 1s ~ 2s