EDAN iM80 User Manual - page 94
Patient Monitor User Manual Monitoring Quick TEMP
- 83 -
and then take it out for starting a new measurement.
2
To ensure optimal accuracy, always confirm that the correct measurement position is
selected.
14.3.2 Measurements for Rectal Temperatures
1. Ensure that the rectal probe (red probe) and probe well are installed.
2. Remove the red probe from the probe well.
3. Observe the Rectal Mode indicator on the display (flashing lower body icon).
4. Load a probe cover. Apply lubricant if desired.
5. Separate the buttocks and gently insert the probe only 1.5cm (5/8 inch), less for infants and
children.
Measuring Position in Rectal
WARNING
Incorrect insertion can cause bowel perforation.
14.3.3 Measurements for Axillary Temperatures
1. Ensure that the white probe and probe well are installed.
2. Remove the probe from the probe well.
3. Press Quick TEMP Setup > Measure Pos and select the Axillary mode.
4.Observe the axillary mode indicator on the display (flashing axillary icon).
5. Load a probe cover.
6. Adjust clothing to visualize the axilla.
7. Avoid folds in the axilla and place the probe tip vertically as high as you can as shown.
8. Place the arm at the patient’s side. Hold in this position without movement of the arm or probe
during the measurement cycle.