EDAN iM80 User Manual - page 158
Patient Monitor User Manual Product Specification
- 147 -
Minimum input slew rate
>2.5V/S
Heart rate
Range
ADU: 15 bpm ~ 300 bpm
PED/NEO: 15 bpm ~ 350 bpm
Accuracy
±1% or 1 bpm, whichever is greater
Resolution
1 bpm
Sensibility
≥300 μVPP
PVC
Range
ADU: 0~300 PVCs/ min
PED/NEO: 0~350 PVCs/ min
Resolution
1 PVCs/min
ST value
Range
-2.0 mV ~ +2.0 mV
Accuracy
The max. of ±0.02 mV or 10% (-0.8 mV ~ +0.8 mV),
whichever is greater.
Resolution
0.01 mV
HR averaging method
Method 1
Normally, heart rate is computed by excluding the
minimum and maximum values from the 12 most recent RR
intervals and averaging the residual 10 RR intervals.
Method 2
If each of three consecutive RR intervals is greater than
1200ms, then the four most recent RR intervals are
averaged to compute the HR.
Range of Sinus and SV Rhythm
Tachy
ADU: 120 bpm ~ 300 bpm
PED/NEO: 160 bpm ~ 350 bpm
Normal
ADU: 41 bpm ~ 119 bpm
PED/NEO: 61 bpm ~159 bpm
Brady
ADU: 15 bpm ~ 40 bpm
PED/NEO: 15 bpm ~ 60 bpm
Range of Ventricular Rhythm
Ventricular Tachycardia
The interval of 5 consecutive ventricular wave is less than
600 ms