EDAN iM80 User Manual - page 47
Patient Monitor User Manual Alarm Information
- 36 -
C.O. Comm Fail
C.O. module failure or
communication failure
High
Stop measuring of
C.O.
module, or
notify biomedical
engineer or
Manufacturer’s
service staff.
C.O. TI No Sensor
C.O.
TI sensor not
connected
Low
Insert injectate
temperature sensor.
C.O. TB No Sensor
C.O.
TB sensor not
connected
Low
Insert TB sensor.
TEMP Out Of Range
TB measuring value is
above measuring range.
High
Please check TB
sensor.
C.O. Lack Param
C.O.
measuring needs
parameters
High
Please input patient’s
height and weight.
AA Out Of Range
The AA concentration
exceeds the accuracy range
of AG module.
High
Reduce AA
concentration.
O
2
Sensor Error
The oxygen sensor of the
sidestream AG module has
a failure.
Medium
Stop measuring of
AG
module, and
notify biomedical
engineer or
Manufacturer’s
service staff.
AG Comm Fail
AG
module failure or
communication failure.
High
Stop measuring
function of AG
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
AG Zero Required
AG module requires zero.
Low
Please perform
zeroing.
AG Self-Testing
AG module is self testing.
Low
Please wait the self
testing finishing.
AG Replace O
2
Sensor
O
2
sensor needs to be
replaced.
High
Please replace the O
2
sensor.
AG Check Adapter
AG module checks adapter.
Low
Please wait check
finishing.
O
2
Cali Required
O
2
needs to be calibrated.
Low
Please calibrate O
2
.