Patient Monitor User Manual Managing Patients
- 48 -
WARNING
1
Changing the patient category may change the arrhythmia and NIBP alarm limits.
Always check alarm limits to make sure that they are appropriate for your patient.
2
For paced patients, you must set Paced to On. If it is incorrectly set to Off, the monitor
could mistake a pace pulse for a QRS and fail to give an alarm during asystole.
6.2 Quick Admit
If you do not have the time or information to fully admit a patient. Complete the rest of the
patient information later. To quickly admit a patient, please:
1.
Select Menu > Patient Setup > Quick Admit, then a message is displayed to ask the user to
confirm to update patient.
2.
Click on No to cancel this operation; click on Yes to continue and the Patient Info window
is displayed, choose Type and Pace and set them to the correct mode.
3.
Select Exit.
6.3 Editing Patient Information
To edit the patient information after a patient has been admitted, select Menu > Patient Setup >
Patient Info., and make the required changes on the popup interface.
6.4 Updating a Patient
You should always perform an update before starting monitoring for a new patient. When you
select Menu > Patient Setup > Quick Admit, or Menu > Patient Setup > New Patient,a
message of Press ‘Yes’ to create new patient profile by clearing all current patient data... is
displayed.
If the user selects Yes, the monitor will update the patient information.
If the user selects No, the monitor won’t update the patient information and returns to
patient setup interface.
NOTE:
Discharging patient will clear the history data in the monitor associated with the patient.
6.5 Central Monitoring System
The monitor can be connected to the central monitoring system. Through the network:
1. The monitor sends patient information, real-time monitoring or measurement data to the central monitoring
system.
2. The real-time monitoring information is displayed on the central monitoring system as the same to the
monitor, and the central monitoring system can perform some bilateral control. For example: changing
patient information, alarm limit and so forth.
For detailed information, please refer to MFM-CMS Central Monitoring System User Manual and CMS Central
Monitoring System User Manual
.
And the monitor supports HL 7 protocol.