EDAN iM80 User Manual - page 51
Patient Monitor User Manual Alarm Information
- 40 -
Probe data error
Offline: NTC resistance >R
0
°C; Short: NTC resistance
°C. Med Put the sensor into the sensor bracket, take it out and measure again. If the problem persists, stop using measuring function of TEMP module, and notify biomedical engineer or manufacturer’s service staff. Probe heater error Single failure Med Put the sensor into the sensor bracket, take it out and measure again. If the problem persists, stop using measuring function of TEMP module, or notify biomedical engineer or manufacturer’s service staff. Probe temp too high The original temperature of sensor >+33 °C & ≤+40 °C. Low Put the sensor into the sensor bracket, measure again after the sensor temperature reaches normal value. Temp Sensor Off After the sensor temperature reaches Predict value, it descends to the value lower than Predict value. Med Reconnect the sensor and make sure that the cable is properly connected. Battery Low Battery Low High Please change the battery or charging. Recorder Out Of Paper Recorder Out Of Paper Low Please install the paper Recorder setup needed The user presses the RECORD button when the monitor is not installed with a recorder. Low Notify the manufacturer’s service staff to install and set the recorder.