Patient Monitor User Manual Alarms
- 24 -
The user can set the audio alarm pause to 60 s, 120 s, or 180 s based on the requirement.
4.3.3 Alarm Mute
To mute the alarm, please select Menu > Maintenace > User Maintain > Alarm Setup,and set
Mute to On, thenpress the
key on the front panel for more than three seconds or press
the Silence shortcut key on the screen directly.
And alarm reminder tone DO is heard every three minutes.
4.3.4 Controlling Alarm Volume
The monitor provides five levels of alarm volume: 1, 2, 3, 4 and 5. For adjusting the alarm
volume, please refer to Section Adjusting Alarm Volume.
4.3.5 Setting Alarm Limits
WARNING
1. Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
2. Setting alarm limits to extreme values may cause the alarm system to become
ineffective.
To change individual measurement alarm limits, take HR alarm for example, please refer to the
following steps:
1. Select the HR Parameter area.
2. Select HR Setup > Alarm Setup. Then set the alarm limit to the desired value from the
popup interface. For how to set the alarm limit, please refer to the following figure:
High
Alarm Limit
Low
Alarm Limit
Upper arrow or
lower arrow to
increase or
decrease the alarm
limit
Setting value of
high alarm limit
Setting value of
low alarm limit