EDAN iM80 User Manual - page 75
Patient Monitor User Manual Monitoring ECG
- 64 -
COUPLET
2 consecutive PVCs
BIGEMINY
Vent Bigeminy
TRIGEMINY
Vent Trigeminy
R ON T
A type of single PVC under the condition that HR
interval is less than 1/3 the average interval, followed by a
compensating pause of 1.25X the average R-R interval (the
next R wave advances onto the previous T wave).
PVC
Single PVCs not belonging to the type of above mentioned
PVCs.
TACHY
5 consecutive QRS complex, RR interval is less than 0.5s.
BRADY
5 consecutive QRS complex, RR interval is longer than 1.5s.
MISSED
BEATS
When HR is less than 100 beats/min., no heart beat is tested
during the period 1.75 times of the average RR interval; or
When HR is higher than 100beat/min, no beat is tested within 1
second.
IRR
The patient has irregular heart rate, check patient’s condition,
electrodes, cables and leads
PNC
After the pacemaker is paced, QRS complex can not be
detected during 300ms.
PNP
After the QRS complex, no pace is detected during 1.75 times
of RR interval.
VBRADY
The patient has irregular HR, and his average HR is less than
60bpm. Check his condition, electrodes, cables and leads.
VENT
The patient has irregular heart rate, check patient’s condition,
electrodes, cables and leads.
8.11.2 ARR Analysis Menu
8.11.2.1 Switching ARR Analysis On and Off
To switch ARR Analysis on or off, in the ECG Setup menu, select ARR Analysis to toggle
between On and Off from the popup interface.
8.11.2.2 PVCs Alarm
Select On in the menu to enable prompt message when an alarm occurs; select Off to disable the
alarm function, and there will be a symbol
beside PVCs.