EDAN iM80 User Manual - page 166
Patient Monitor User Manual Product Specification
- 155 -
MAP: 20 mmHg ~ 110 mmHg
Cuff pressure measuring range
0 mmHg ~ 300 mmHg
Pressure resolution
1mmHg
Maximum mean error
±5mmHg
Maximum standard deviation
8mmHg
Maximum measuring period
Adult/Pediatric
120s
Neonate
90s
Typical measuring period
30s ~ 45s (depend on HR/motion disturbance)
Overpressure protection (Dual overpressure protection)
Adult
297±3mmHg
Pediatric
240±3mmHg
Neonatal
147±3mmHg
PR
Measuring range
40 bpm ~240bpm
Accuracy
±3bpm or 3.5%, whichever is greater
A.10.2 NIBP from M3600 Module
Method
Oscillometric
Mode
Manual, Auto, Continuous
Measuring Interval in AUTO Mode
1/2/3/4/5/10/15/30/60/90 min, 2/4/8h
PR Range
Adult/ Pediatric mode: 40bpm ~ 200bpm
Neonatal mode: 40 bpm ~ 240bpm
PR Accuracy
± 2 bpm or 2% of the readings
Measuring Type
SYS, DIA, MAP
Measuring Range
Adult/ Pediatric Mode
SYS: 60 mmHg ~ 250 mmHg
DIA: 40 mmHg ~ 200 mmHg
MAP: 45 mmHg ~ 235 mmHg
Neonatal Mode
SYS: 40 mmHg ~ 120 mmHg
DIA: 20 mmHg ~ 90 mmHg
MAP: 30 mmHg ~ 100 mmHg