EDAN iM80 User Manual - page 173
Patient Monitor User Manual Product Specification
- 162 -
Response Time
Accuracy(Standard Conditions)
GAS
Measurement Range
Accuracy
CO
2
0 to 15 vol%
15 to 25 vol%
±(0.2 vol% + 2% of reading)
Unspecified
N
2
O
0 to 100 vol%
±(2 vol% + 2% of reading)
HAL, ENF, ISO
0 to 8 vol %
8 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
SEV
0 to 10 vol %
10 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
DES
0 to 22 vol %
22 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
O
2
0 to 100 vol %
±(1 vol% + 2% of reading)
Accuracy (All Conditions)
Gas
Accuracy
CO
2
±(0.3kPa + 4% of reading)
N
2
O
±(2kPa + 5% of reading)
Agents
±(0.2kPa + 10% of reading)
O
2
±(2kPa + 2 of reading)
Apnea Alarm Delay 20s~60s
Item
Operating conditions
Storage conditions
Temperature(
°C)
5 to 50
-40 ~ 70
Relative humidity
5 ~ 100% RH(Condensing )
Atmospheric
pressure
525 ~ 1200hPa
200 ~ 1200hPa
Interfering gas and vapor effects
Gas or vapour
Gas level
CO
2
Agents
N
2
O
ISA CO
2
ISA AX+
N
2
O
4)
60 vol%
_ 2)
_ 1)
_ 1)
_ 1)
HAL
4)
4 vol%
_ 1)
_ 1)
_ 1)
_ 1)
ENF, ISO, SEV
4)
5 vol%
+8% of
_ 1)
_ 1)
_ 1)