EDAN iM80 User Manual - page 48
Patient Monitor User Manual Alarm Information
- 37 -
AG Software Error
AG module software
abnormal
High
Please replace
software revision.
AG Hardware Error
AG module has hardware
failure.
High
Please check whether
the hardwares work
properly.
AG Motor Error
AG module motor abnormal
High
Please check whether
the motor works
properly.
AG Uncalibrated
AG module uncalibrated
Low
Please calibrate the
AG module.
AG Replace Adapter
AG module needs to change
adapter.
High
Please replace the
adapter.
O
2
Out Of Range
O
2
is out of range.
High
Please make O
2
range
resume normal.
AG TEMP Out Of
Range
AG module temperature out
of range
High
Please make the
temperature resume
normal.
AG Baro Press Out Of
Range
AG module baro pressure
out of pressure
High
Please make the baro
pressure
value
resume normal.
AG AA Id Unreliable
AG module can’t identify
the AG agent.
Medium
Reduce gas agent
type.
AG Span Calib In
Progress
AG module is calibrating.
Low
Please wait
calibration finishing.
AG Calibration Fail
AG module calibration
failure
Medium
Please check whether
the module works
properly.
Unable To Calibrate
AG module can’t be
calibrated.
Medium
Please check whether
the module works
properly.
AG Zero In Progress
AG module is zeroing.
Low
Please wait zeroing.
AG Occlusion
The sampling line is
clogged.
Medium
Replace the sampling
line.
AG Init Fail
AG module has a failure.
High
AG module works
improperly.
AG Data Limit Error
AG module has a failure.
High
AG module works
improperly.