EDAN iM80 User Manual - page 157
Patient Monitor User Manual Product Specification
- 146 -
Notch
50Hz/60Hz (Notch filter can be turned on or off manually)
Differential Input Impendance
>5MΩ
Input Signal Range
±8mV PP
Accuracy of Input Signal
Reconstruction
The total error and frequency response comply with
ANSI/AAMI EC13:2002, Sect. 4.2.9.8.
Electrode Offset Potential
Tolerance
±500mV
Auxiliary Current (Leads off
detection)
Active electrode:
Reference electrode:
Input Offset Current
≤0.1μA
Recovery time after
Defibrillation
Leakage current of patient
Scale signal
1mVPP, accuracy is ±5%
System noise
ESU Protection
Incision mode: 300W
Congelation mode: 100W
Restore time:
≤10s
Meets the requirements of ANSI/AAMI EC13-2002: Sect.
4.1.2.1 a)
Noise Suppression of
Electrotome
Tested according to the test method in EC13: 2002
Sect.5.2.9.14, it accords with the standard.
Pace Pulse
Pulse indicator
Pulse is marked if the requirements of ANSI/AAMI
EC13:2002, Sect. 4.1.4.1 are met:
Amplitude: ±2 mV ~ ±700 mV
Width: 0.1 ms ~2 ms
Ascending time: 10
μs ~ 100 μs
Pulse Rejection
Pulse is rejected if the requirements of ANSI/AAMI
EC13-2002: Sect. 4.1.4.1 are met:
Amplitude: ±2 mV ~ ±700 mV
Width: 0.1 ms ~2 ms
Ascending time: 10
μs ~100 μs