AwRR
0 rpm ~ 150 rpm
Resolution
EtCO
2
1mmHg
FiCO
2
1mmHg
AwRR
1 rpm
EtCO
2
Accuracy
± 2 mmHg, 0 to 40 mmHg
± 5 % of reading, 41 to 70 mmHg
± 8 % of reading, 71 to 100 mmHg
± 10 % of reading, 101 to 150 mmHg
AwRR Accuracy
± 1 rpm
Suffocation Alarm Delay
10s, 15s, 20s, 25s, 30s, 35s, 40s, default value is 20
second.
Sample Gas Flowrate
50ml/min
Stability
Short Term Drift
Drift over 4 hours
Long Term Drift
120 hour period
O
2
Compensation
Range
0 ~ 100%
Resolution
1%
Default
16%
Initialization time
It displays the value within 15s and meets the requirement
for measurement accuracy within 2min. (Mainstream)
It displays the value within 20s and meets the requirement
for measurement accuracy within 2min. (Sidestream)
Response time
60ms (Mainstream)
Calibration
Not required.
Barometric
pressure
compensation
User setup
Apnea Alarm Delay
10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s; default value is 20s.
Interfering Gas and Vapor Effect on EtCO
2
Measurement Values:
Gas or vapor
Gas level (%)
Quantitative effect/Comments
Nitrous oxide
Halothane
60
4
Dry and Saturated Gas
0 – 40 mmHg: ± 1 mmHg additional error