EDAN iM80 User Manual - page 169
Patient Monitor User Manual Product Specification
- 158 -
A.13 Quick TEMP
Measuring Range
25
°C ~ 45°C (77 º F ~113 º F)
Operating Temp
10
°C ~ 40°C (50 º F ~104 º F)
Sensor Type
Oral/Axillary sensor, Rectal sensor
Resolution
0.1
°C (0.1 º F)
Accuracy
±0.1°C (25°C ~ 45°C) or
± 0.2℉ (77 º F ~113 º F)
Response time
Update time
1s ~ 2s
A.14 IBP
Pressure measuring range
-50 to +300 mmHg
Resolution
1 mmHg
Accuracy (without sensor)
± 2 % or ±1 mmHg, whichever is greater
Pressure sensor
Sensitivity
5 (μV/V/mmHg)
Impedance
300 to
3000 Ω
Frequency response
d.c. to 12.5 or 40 Hz
Zero
Range:
±200 mmHg
Accuracy: 1 mmHg
Measuring range
Art
0 mmHg to 300 mmHg
PA
-6 to +120mmHg
CVP/RAP/LAP/ICP
-10 to +40 mmHg
P1/P2
-50 to +300 mmHg
Volume displacement of MSI
4.5 x 10
-4
in
3
/ 100 mmHg
A.15 CO
2
Method
Infra-red Absorption Technique
Unit
mmHg, %, Kpa
Measuring Range
EtCO
2
0 mmHg ~ 150 mmHg
FiCO
2
3 mmHg ~50 mmHg