EDAN iM80 User Manual - page 161
Patient Monitor User Manual Product Specification
- 150 -
Monitor: 0.5Hz to 40Hz
Surgery: 1Hz to 20Hz
CMRR (Common Mode
Rejection Ratio)
Diagnosis: >95dB (the Notch filter is off)
Monitor: >105dB (the Notch filter is on)
Surgery: >105dB (the Notch filter is on)
Notch
50Hz/60Hz (Notch filter can be turned on or off manually)
Differential Input
Impendance
>5MΩ
Input Signal Range
±8mV PP
Accuracy of Input Signal
Reconstruction
The total error and frequency response comply with ANSI/AAMI
EC13:2002, Sect. 4.2.9.8.
Electrode Offset Potential
Tolerance
±500mV
Auxiliary Current (Leads
off detection)
Active electrode:
Reference electrode:
Input Offset Current
≤0.1μA
Recovery time after
Defibrillation
Leakage current of patient
Scale signal
1mVPP, accuracy is ±5%
System noise
ESU Protection
Incision mode: 300W
Congelation mode: 100W
Restore time:
≤10s
Meets the requirements of ANSI/AAMI EC13-2002: Sect. 4.1.2.1
a)
Noise Suppression of
Electrotome
Tested according to the test method in EC13: 2002 Sect.5.2.9.14,
it accords with the standard.
Pace Pulse
Pulse indicator
PACE pulses are marked by the PACE indicator if the following
conditions are met:
Amplitude: ±2 mV ~ ±700 mV (12-lead)