Patient Monitor User Manual Alarm Information
- 35 -
NIBP Pulse Abnormal
Abnormal oscillometric
waveform
Low
Retry twice. Check
for hyperkinesia or
arrhythmia.
NIBP Pulse Signal
Weak
Pulse is too low to measure.
Low
Check the patient’s
condition or the
wrapped cuff.
TEMP T1 Sensor Off
Temperature cable of TEMP
channe1 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected
TEMP T2 Sensor Off
Temperature cable of TEMP
channe2 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
TEMP Comm Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
YY Sensor Off (YY
stands for the IBP label
name)
IBP sensor falls off.
Low
Check the sensor
connection and
reconnect the sensor.
YY Comm Fail (YY
stands for the label
name)
IBP module failure or
communication failure
High
Stop measuring
function of IBP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.