Patient Monitor User Manual Alarm Information
- 32 -
SpO
2
Sensor Off
SpO
2
sensor may be
disconnected from the
patient or the monitor.
Low
Make sure the sensor
is well connected to
the patient’s finger or
other parts.
SpO
2
No Sensor
SpO
2
sensor was not
connected well or
connected to the monitor, or
the connection is loose.
Low
Make sure the
monitor and sensor is
well connected,
reconnect the sensor.
SpO
2
Comm Fail
SpO
2
module failure or
communication failure
High
Stop using measuring
function of SpO
2
module, and notify
biomedical engineer
or
manufacturer’s
service staff.
SpO
2
Low Perfusion
The pulse signal is too
weak or the perfusion of
the measurement site is too
low
Low
Reconnect the SpO
2
sensor and change
the measurement site.
If problem exists,
please
notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Comm Fail
NIBP module failure or
communication failure
High
Stop using measuring
function of NIBP
module, and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Loose Cuff
Cuff is no properly wrapped
or no cuff exists.
Low
Properly wrap the
cuff.
NIBP Excessive
Pressure
Pressure has exceeded the
specified upper safety limit.
Low
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify biomedical
engineer or
manufacturer’s
service staff.