EDAN iM80 User Manual - page 49
Patient Monitor User Manual Alarm Information
- 38 -
AG Usa Error
AG module has a failure.
High
AG module works
improperly.
AG Cal Fail
AG module fails to
calibrate.
High
AG module works
improperly.
AG Zref Fail
AG module fails to zero.
High
AG module works
improperly.
AG Change Oxygen
Sensor
Replace oxygen sensor of
AG module.
Low
Please wait changing
finishing.
AG No Oxygen
Sensor
The oxygen sensor falls off
from the AG module.
High
Connect the sensor
again.
AG Mixed Agents
AG module detects mixture
gas agent.
Medium
Close the subsidiary
gas agent.
CO
2
Occlude
Water trap of SideStream is
occluded.
Low
Make sure the gas
exhaust works well
CO
₂ Out Of Range
The
CO
2
concentration
exceeds the accuracy range
of AG module.
High
Reduce CO
2
concentration.
CO
2
Sensor Faulty
CO
2
module failure
High
Stop measuring
function of CO
2
module, notify
biomedical engineer.
CO
2
Sensor Over
Temp
CO
2
Comm Fail
CO
2
module failure or
communication failure
High
Stop using CO
2
alarm
function, and notify
biomedical engineer
or Manufacturer’s
service staff.
CO
2
Ram Error
CO
2
module failure
High
CO
2
Rom Error
CO
2
module failure
High
CO
2
Zero Required
Zero calibration failure
Low