Patient Monitor User Manual Basic Operation
- 21 -
3.9 Understanding Screens
Your monitor comes with a set of preconfigured screens, optimized for common monitoring
scenarios such as OR adult, or ICU neonatal. A screen defines the overall selection, size and
position of waves, numerics and shortcut keys on the monitor screen when you switch on the
monitor. You can easily switch between different screens during monitoring. Screens do NOT
affect alarm settings, patient category and so forth.When you switch from a complex to a less
complex screen layout, some measurements may not be visible but are still monitored in the
background. For detailed information, please refer to Chapter User Interface.
3.10 Calibrating Screens
To calibrate the screen, please refer to the following steps:
1. Select the Touch Calib shortcut key on the screen directly or select Menu > Maintenance >
User Maintain, then input maintenance password ABC, then select TouchScr Calibration
from the popup interface.
2. The symbol
appears on the screen.
3. Click on the central point of the symbol
.
4. After successful calibration, the message of Screen Calibration Completed appears on the
screen. Then select Exit to finish the calibration.
3.11 Disabling the TouchScreen
The user can disable touchscreen operation by pressing and
holding the Menu shortcut key for 3
seconds. A message of Screen Locked and the symbol
are displayed at the bottom of screen
if the touchscreen is disabled. To enable the touchscreen operation, turn the knob to select the
symbol
and press it.
3.12 Using the Barcode Scanner
To enter the barcode setup menu, please select Menu > Maintenance > User Maintain, after
entering the required password ABC, select Other Setup > BarCode Setup. Then the user can
set serial No, last Name, first Name and so on.
If Auto Update is set to On, the patient information is updated automatically by using a bar code
scanner.If Auto Update is set to Off, the user needs to update the patient information manually.