Patient Monitor User Manual Product Specification
- 164 -
Parameters
Sevoflurane(SEV), Desflurane(DES), awRR, MAC
Measurement
Principle
CO
2
, N
2
O, anaesthesia agent: infra-red absorption characteristic
Warm-up Time
Concentrations are reported and the automatic agent indentification is
running within 10 seconds.
20 seconds for IRMA AX+.
Rise Time
CO
2
≤ 90ms
O
2
≤ 300ms
N
2
O
≤ 300ms
HAL, ISO, ENF, SEV, DES
≤ 300ms
Primary Agent
Threshold
0.15 vol%
Secondary Agent
Threshold
0.2 vol% + 10% of total agent concentration
Agent
Identificaiton
Time
Response Time
Accuracy(Standard Conditions)
Gas
Range
Accuracy
CO
2
0 ~ 10 vol%
10 ~ 15vol%
15 ~ 25 vol%
±(0.2 vol% + 2% of reading)
±(0.3 vol% + 2% of reading)
Unspecified
N
2
O
0 to 100%
±(2 vol% + 2% of reading)
HAL
ISO
ENF
0 to 8
8 to 25
±(0.15 vol% + 5% of reading)
Unspecified
SEV
0 to 10
10 to 25
±(0.15 vol% + 5% of reading)
Unspecified
DES
0 to 22
22 to 25
±(0.15 vol% + 5% of reading)
Unspecified
Accuracy (All Conditions)
GAS
Accuracy