EDAN iM80 User Manual - page 162
Patient Monitor User Manual Product Specification
- 151 -
Width: 0.1 ms ~2 ms
Ascending time: 10
μs ~ 100 μs
Pulse Rejection
Pulse is rejected if the requirements of ANSI/AAMI EC13-2002:
Sect. 4.1.4.1 are met:
Amplitude: ±2 mV ~ ±700 mV
Width: 0.1 ms ~2 ms
Ascending time: 10
μs ~100 μs
Minimum input slew rate
>2.5V/S
Heart rate
Alarm Range
ADU: 15 bpm ~ 300 bpm
PED/NEO: 15 bpm ~ 350 bpm
Accuracy
±1% or 1 bpm, whichever is greater
Resolution
1 bpm
Sensibility
≥300 μVPP
PVC
Range
ADU: 0~300 PVCs/ min
PED/NEO: 0~350 PVCs/ min
Resolution
1 PVCs/min
ST value
Range
-2.0 mV ~ +2.0 mV
Accuracy
The max. of ±0.02 mV or 10% (-0.8 mV ~ +0.8 mV), whichever
is greater.
Resolution
0.01 mV
HR averaging method
Method 1
Normally, heart rate is computed by excluding the minimum and
maximum values from the 12 most recent RR intervals and
averaging the residual 10 RR intervals.
Method 2
If each of three consecutive RR intervals is greater than 1200ms,
then the four most recent RR intervals are averaged to compute
the HR.
Range of Sinus and SV Rhythm
Tachy
ADU: 120 bpm ~ 300 bpm
PED/NEO: 160 bpm ~ 350 bpm