EDAN iM80 User Manual - page 76
Patient Monitor User Manual Monitoring ECG
- 65 -
WARNING
When the PVCs Alarm is set to OFF, the monitor won’t give an alarm prompt even if an
alarm occurs. In order to avoid endangering the patient’s life, the user should use this
function cautiously.
8.11.2.3 ARR Relearning
Pick this item to start a learning procedure, and ECG ARR LEARNING is displayed on the
screen. The ECG ARR LEARNING will start automatically in the following status:
Changing lead type;
Connectiong leads;
Updating the patients;
Starting ARR learning manually;
After the ARR analysis is switched on;
The module is set to on;
Calibration mode is changed to nomoral measurement mode;
Exiting the Demo mode;
Exiting the standby mode;
8.11.2.4 ARR Alarm
The users can switch on or off all arrhythmia alarms by selecting ECG Setup > ARR Analysis >
ARR Alarm. And some arrhythmia alarms can be individually switched on or off. They are:
ASYSTOLE, VFIB/VTAC, R-ON-T, VT>2, COUPLET, PVC, BIGEMINY, TRIGEMINY,
TACHY, BRADY, MISSED BEATS, IRR, PNC, PNP, VBRADY and VENT.
To switch individual alarm on or off, select ECG Setup > ARR Analysis > ARR Alarm. The
user can set the individual ARR alarm from the popup interface.
8.12 12-Lead ECG Monitoring
8.12.1 Diagnosis Function
If iM80 is equipped with 12 leads monitoring, the device has automatical diagnosis function. To
perform this function, please:
1
Set Lead Type in the ECG Setup menu to 12 Leads andset Display in the ECG Setup
menu to 12 Leads.
2
Select the ECG Analysis shortcut key on the screen.
3
The Diagnosis Review interface is displayed, shown in figure a. And the diagnosis result is
displayed on the interface after approximately 10 seconds, shown in figure b.