Patient Monitor User Manual Alarm Information
- 27 -
PNC
PACE NOT CAPTURE: After the pacemaker is paced,
QRS complex can not be detected during 300ms.
User-selectable
PNP
PACER NOT PACED: After the QRS complex, no
pace is detected during 1.75 times of RR interval.
User-selectable
VBRADY
VENTRICULAR BRADYCARDIA: The patient has
irregular HR, and his average HR is less than 60bpm.
Check his condition, electrodes, cables and leads.
User-selectable
VENT
VENTRICULAR RHYTHM: The patient has irregular
heart rate, check patient’s condition, electrodes, cables
and leads.
User-selectable
RESP APNEA
RESP can not be measured within specific time
interval.
High
RR High
RESP measuring value is above upper alarm limit.
User-selectable
RR Low
RESP measuring value is below lower alarm limit.
User-selectable
HR High
HR measuring value is above the upper alarm limit.
User-selectable
HR Low
HR measuring value is below the lower alarm limit.
User-selectable
SpO
2
High
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
Low
SpO
2
measuring value is below lower alarm limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too weak, so the
monitor can’t detect the pulse signal.
High
PR High
PR measuring value is above upper alarm limit.
User-selectable
PR Low
PR measuring value is below lower alarm limit.
User-selectable
T1 High
Measuring value of T1 channel is above upper alarm
limit.
User-selectable
T1 Low
Measuring value of T1 channel is below lower alarm
limit.
User-selectable
T2 High
Measuring value of T2 channel is above upper alarm
limit.
User-selectable
T2 Low
Measuring value of T2 channel is below lower alarm
limit.
User-selectable
TD High
Measuring value of TD channel is above upper alarm
limit.
User-selectable
TEMP High
Measuring value of TEMP is above upper alarm limit.
User-selectable
TEMP Low
Measuring value of TEMP is below lower alarm limit.
User-selectable
SYS High
SYS measuring value is above upper alarm limit.
User-selectable
SYS Low
SYS measuring value is below lower alarm limit.
User-selectable
DIA High
DIA measuring value is above upper alarm limit.
User-selectable